Step 7. Move the Mass Spec Bench to its final position
1
When the LC/MS or GC/MS instrument has been placed upon the Mass Spec
Bench, move the Mass Spec Bench to it final installation position.
2
When the Mass Spec Bench is in the desired position, lower the extendable
pads inside the corner castors.
To extend the pads, pull the handle out and turn the ratchet in the
clockwise direction. If the ratchet is not turning in the correct direction,
flip the switch to change the ratchet direction.
Extend the pads so that the castor wheels no longer touch the floor.
3
When the corner castors have been extended, extend the feet on the pump
tray.
This action isolates the rough pump vibration from the Mass Spec Bench.
Note that access to the rear castors can be difficult. Use a 3/4 wrench to
extend the feet of the pump tray so that the pump tray is no longer in
contact with the Mass Spec Bench. The pump tray needs only 1/8-inch of
height above the support rails.
Figure 22
Extendable Rubber Feet
The ratchet handle must be
pulled out in order to extend
the pads.
Содержание Agilent G3215A
Страница 1: ...Agilent Technologies Agilent G3215A Mass Spec Bench User Guide Revision A May 2011...
Страница 4: ...4 Mass Spec Bench User Guide...
Страница 25: ...Installation 2 Mass Spec Bench User Guide 25 Figure 15 Single pump orientation Figure 16 Dual pump orientation...
Страница 30: ...2 Installation 30 Mass Spec Bench User Guide Figure 21 Extending the pump tray feet...
Страница 35: ...Operation 3 Mass Spec Bench User Guide 35 Figure 26 Fan power switch...
Страница 41: ...Troubleshooting 4 Mass Spec Bench User Guide 41...