3-11
Measurements Applications
Measuring Laser Drift
laser line of interest may have since drifted to a greater
value. Note that the minimum wavelength and minimum
power may not have occurred simultaneously.
Display shows the
total
drift from the reference since the
drift measurement was started. Values represent the
minimum wavelength and power drift values subtracted
from the maximum drift values.
5
In the
List by WL
and
List by Power
displays, use the
and
softkeys to view
the reference values (wavelength and power values of each laser line before the
test was started).
During the measurement, you can change the display mode to
Peak WL
,
List by
WL
,
List by Power
, or
Avg WL
. When
List by WL
or
List by Power
is selected, the signal
list is sorted by reference values and not by the current, maximum, or minimum
values.
To restart the drift measurements, press
RESET
. This resets the reference val-
ues.
Содержание Agilent 86120B
Страница 1: ...Agilent 86120B Multi Wavelength Meter User s Guide ...
Страница 11: ...Contents 3 Contents Power Cords 8 16 Agilent Technologies Service Offices 8 18 ...
Страница 12: ......
Страница 38: ......
Страница 68: ......
Страница 100: ...4 18 Programming Monitoring the Instrument ...
Страница 132: ......
Страница 162: ...5 30 Programming Commands CALCulate1 Subsystem Query Response For normal update 34123 For fast update 4268 ...
Страница 230: ......
Страница 231: ...7 Definition of Terms 7 3 Specifications 7 6 Regulatory Information 7 10 Specifications and Regulatory Information ...
Страница 241: ...7 11 Specifications and Regulatory Information Regulatory Information Declaration of Conformity ...
Страница 242: ...7 12 Specifications and Regulatory Information Regulatory Information Front view of instrument Rear view of instrument ...
Страница 248: ...8 6 Reference Menu Maps Delta On Menu Delta Off Menu ...
Страница 250: ...8 8 Reference Menu Maps System Setup Menu ...
Страница 268: ......