3-6
Measurements Applications
Measuring Signal-to-Noise Ratios
4
To select the wavelength reference for measuring the noise, do the following
steps:
a
Press
WL REF
, and
• press
AUTO
to let the instrument interpolate the wavelength,
or
• press
USER
to select the last wavelength manually entered.
b
If you chose
USER
, you can specify the wavelength by pressing
USER WL
. Use
the
softkey to select the digit that requires editing. Use the
and
softkeys to change the value.
c
Press
RETURN
.
5
While the signal-to-noise measurements are displayed, you can press
PEAK
anytime to select the signal with the highest power.
Содержание Agilent 86120B
Страница 1: ...Agilent 86120B Multi Wavelength Meter User s Guide ...
Страница 11: ...Contents 3 Contents Power Cords 8 16 Agilent Technologies Service Offices 8 18 ...
Страница 12: ......
Страница 38: ......
Страница 68: ......
Страница 100: ...4 18 Programming Monitoring the Instrument ...
Страница 132: ......
Страница 162: ...5 30 Programming Commands CALCulate1 Subsystem Query Response For normal update 34123 For fast update 4268 ...
Страница 230: ......
Страница 231: ...7 Definition of Terms 7 3 Specifications 7 6 Regulatory Information 7 10 Specifications and Regulatory Information ...
Страница 241: ...7 11 Specifications and Regulatory Information Regulatory Information Declaration of Conformity ...
Страница 242: ...7 12 Specifications and Regulatory Information Regulatory Information Front view of instrument Rear view of instrument ...
Страница 248: ...8 6 Reference Menu Maps Delta On Menu Delta Off Menu ...
Страница 250: ...8 8 Reference Menu Maps System Setup Menu ...
Страница 268: ......