28
User Manual
4
Instrument Connection
Analog Input (Signal) Connection
The signal must be within the range of -18 millivolts to 1 volt. A simple voltage
divider resistor network may be used to attenuate larger signals to this range.
Each input has three pins to which the high (+) and low (–) sides of the signal
and the shield are connected.
Use an instrument-specific cable (Table 6) whenever possible. Use a general
purpose cable in those cases where a custom cable is not available.
Appendix A lists pin assignments for all open cables.
1
Turn off the power to the instrument supplying the analog output signal.
2
For instrument-specific cables–
Connect one end of the cable to the
instrument's analog output using the instrument manufacturer's
instructions.
For general purpose cables –
Use the pin assignments shown in
Appendix A, “Cables, and the instrument manufacturer's instructions to
connect the two signal wires and ground to the instrument.
3
Connect the cable to the appropriate (A or B) analog input on the rear
panel of the 35900 (Figure 8).
4
If needed, repeat for the other channel.
C A U T I O N
For correct operation of the 35900, the input voltage must be in the range of -18 me to
+1 V. Voltages below –10 V or above +10 V may damage the instrument.
Содержание Agilent 35900E
Страница 1: ...Agilent Technologies Agilent 35900E Dual Channel Interface User Manual ...
Страница 6: ...6 User Information Manual ...
Страница 36: ...36 User Manual 4 Instrument Connection ...
Страница 48: ...48 User Manual 6 Parts Replacement ...
Страница 51: ...Instrument Parts Breakdown 7 User Manual 51 Figure 15 Parts breakdown 1 2 3 4 5 ...
Страница 53: ...Instrument Parts Breakdown 7 User Manual 53 Figure 16 Parts breakdown inside the 35900 1 2 3 4 ...
Страница 54: ...54 User Manual 7 Instrument Parts Breakdown ...
Страница 60: ...60 User Manual A Cables ...