24
User Manual
3
Computer Connection
Configuration for RS-232 operation
The GPIB/RS-232 MIO card contains a single 8 position configuration switch
(Figure 6). When the position 1 switch is set to ON, the MIO card uses the
RS-232 communication mode. In this mode, the ON switch positions for
switches 2 through 8 are defined in Table 3.
An example setting for this switch is shown in Figure 6.
Table 3
RS-232 communication mode
Switch
ON Action
2
Handshake enabled
3
Hardware handshake (otherwise software handshake)
4
Parity enable
5
Even parity (otherwise odd parity)
6, 7, 8
Baud rate (see Table 4)
Figure 6
Example settings for RS-232
1
8
On
Off
Rear