User Manual
5
Startup
Pre-Run Checklist
Before using the 35900 with the operating software on the host computer,
make sure you have completed the following:
❏
Rear Panel configuration switches 3 through 6 are DOWN (normal
operation)
❏
MIO card is correctly configured for host computer communications
❏
MIO card connected to host computer or network
❏
Analog signal cables connected
❏
Remote control cables connected (if needed)
❏
Digital I/O cables connected (if needed)
❏
Power cord connected
35900 Power Up
Power up the 35900 by pressing the line switch (lower left corner of the front
panel).
•
All front panel indicators should turn on for 1 to 2 seconds.
•
All indicators should turn off, except for POWER (on steadily) and FAULT
(blinks while self-tests are performed).
•
After approximately 30 seconds, the FAULT indicator should turn off. Some
other indicators may turn on, depending on the instrument configuration.
WA RN I N G
The power cord must be connected to an outlet with a protective earth contact to
ground the unit and minimize shock hazard.