4-32
Performance Tests
8. Test Port Crosstalk
In Case of
Difficulty
W A R N I N G
These servicing instructions are for use by qualified personnel only.
To avoid electrical shock, do not perform any servicing unless you are
qualified to do so.
W A R N I N G
The opening of covers or removal of parts is likely to expose
dangerous voltages. Disconnect the instrument from all voltage
sources while it is being opened.
W A R N I N G
The power cord is connected to internal capacitors that may remain
live for five seconds after disconnecting the plug from its power
supply.
1
Remove the instrument top cover. Using an 8 lb-inch torque wrench, verify that
all
semi-rigid cables connected to the sampler/mixer assemblies are tight. In
addition tighten any loose screws on the sampler/mixer assemblies (A4/5/6)
and the pulse generator assembly (A7).
2
Remove the instrument bottom cover. Refer to
. Verify that cables
W1, W31 and W32 are tight.
3
Repeat this test.