![Agilent Technologies 8702D Скачать руководство пользователя страница 232](http://html.mh-extra.com/html/agilent-technologies/8702d/8702d_installation-manual_2867953232.webp)
6-66
Performance Tests – Option 011
17. Receiver Magnitude Frequency Response (Option 002 Only)
for the receiver input A. Press
MEAS
,
INPUT PORTS
,
A
Step 6
.
36
through
to measure the receiver worst case 3nd
harmonic for the receiver input A.
In Case of
Difficulty
W A R N I N G
These servicing instructions are for use by qualified personnel only.
To avoid electrical shock, do not perform any servicing unless you are
qualified to do so.
W A R N I N G
The opening of covers or removal of parts is likely to expose
dangerous voltages. Disconnect the instrument from all voltage
sources while it is being opened.
W A R N I N G
The power cord is connected to internal capacitors that may remain
live for five seconds after disconnecting the plug from its power
supply.
1
If source harmonics fail, replace the A3 source assembly.
2
If the A receiver harmonics fail, replace the A5 sampler/mixer assembly.
3
Replace the A6 sampler/mixer assembly if the B receiver harmonics fail.
17. Receiver Magnitude Frequency Response
(Option 002 Only)
Perform this test to determine the receiver’s ability to transfer information
from RF to IF and how accurately that information is processed and displayed.
Required
Equipment
External Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 83640A
Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 436A/437B/438A
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 8482A
3-way Power Splitter, 50
Ω
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 11850C
Attenuator, 10 dB . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 8491A Option 010
RF cable set, 50
Ω
, Type-N . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 11851B
Cable, BNC, 24-inch . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-1840
Adapter, Type-N (m) to Type-N (m) . . . . . . . . . . . . . . Agilent P/N 1250-1475