![Agilent Technologies 8702D Скачать руководство пользователя страница 42](http://html.mh-extra.com/html/agilent-technologies/8702d/8702d_installation-manual_2867953042.webp)
2-6
Automated Verification
Step 3. Device Verification
17
Connect the test port cables as shown
.
Figure 2-3. Transmission Calibration Setup
18
Press
TRANSMISSION
,
FWD TRANS THRU
.
19
When the Agilent 8702D finishes the measurement, press
FWD MATCH THRU
.
20
When the Agilent 8702D finishes the measurement, press
REV TRANS THRU
.
21
When the Agilent 8702D finishes the measurement, press
REV MATCH THRU
.
22
When the Agilent 8702D displays,
PRESS ’DONE’ IF FINISHED WITH
STD(s)
, press
STANDARDS DONE
,
DONE 2-PORT CAL
.
23
Press
SAVE/RECALL
,
SELECT DISK
,
INTERNAL MEMORY
,
RETURN
,
SAVE STATE
to save
the calibration into the Agilent 8702D internal memory.
24
When the Agilent 8702D finishes saving the instrument state, press
SELECT
DISK
,
INTERNAL DISK
.
Step 3. Device Verification
1
Press
SYSTEM
,
SERVICE MENU
,
TESTS
,
28
,
x1
,
EXECUTE TEST
.
2
At the prompt, connect the 20 dB attenuator (supplied in the verification kit)
as shown in
.
3
Press
CONTINUE
to run the test:
• If you switched the record function
off
, you have to press
CONTINUE
after