![Agilent Technologies 8702D Скачать руководство пользователя страница 222](http://html.mh-extra.com/html/agilent-technologies/8702d/8702d_installation-manual_2867953222.webp)
6-56
Performance Tests – Option 011
15. Receiver Phase Compression
In Case of
Difficulty
W A R N I N G
These servicing instructions are for use by qualified personnel only.
To avoid electrical shock, do not perform any servicing unless you are
qualified to do so.
W A R N I N G
The opening of covers or removal of parts is likely to expose
dangerous voltages. Disconnect the instrument from all voltage
sources while it is being opened.
W A R N I N G
The power cord is connected to internal capacitors that may remain
live for five seconds after disconnecting the plug from its power
supply.
1
If the analyzer fails the channel A compression test, suspect the A5 A sampler
assembly. Repeat this test. Replace the sampler assembly if the problem still
exists.
2
Suspect the A6 B sampler assembly if the analyzer fails the channel B
compression test. Repeat this test. Exchange the sampler assembly if the
failure still occurs.
15. Receiver Phase Compression
Perform this test to verify the compression/expansion phase relationships of
the analyzer’s receiver samplers.
Required
Equipment
Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 437B/438A
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 8482A
Attenuator, 3 dB . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 8491A Option 003
Attenuator, 10 dB . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 8491A Option 010
Attenuator, 20 dB . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 8491A Option 020
Cable (2), 50
Ω
, Type-N, 24-inch . . . . . . . . . . . . . . . . . . Agilent P/N 8120-4781
2-way Power Splitter (2), 50
Ω
. . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 11667A
Adapter, Type-N (m) to Type-N (m) . . . . . . . . . . . . . . Agilent P/N 1250-1475
Additional Required Equipment for Analyzers with Option 006
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 8481A