![Agilent Technologies 8702D Скачать руководство пользователя страница 109](http://html.mh-extra.com/html/agilent-technologies/8702d/8702d_installation-manual_2867953109.webp)
4-53
Performance Tests
13. Test Port Receiver Magnitude Dynamic Accuracy
Figure 4-32. Test Port 1 Magnitude Dynamic Accuracy Test Setup
18
On the Agilent 8702D:
a
Press
MENU
,
POWER
.
b
Using the analyzer front panel knob, adjust the test port power for a reading
of 0.000 dB on the Agilent 8902A.
c
Wait for the reading to stabilize.