2-15
Making Measurements
Defining Laser-Line Peaks
Defining Laser- Line Peaks
The Agilent 86120C uses two rules to identify valid laser- line peaks.
Understanding these rules is essential to getting the most from your
measurements. For example, these rules allow you to “hide” AM mod-
ulation sidebands or locate laser lines with small amplitudes.
In order to identify a laser line, the laser- line must meet
both
of the
following rules:
• Power must be greater than the power established by the
peak threshold
limit
• Power must rise and then fall by at least the
peak excursion
value
In addition, the inp ut wavelength range can be limited as described in
this section.
Peak threshold
limit
The peak threshold limit is set by subtracting the peak threshold value
from the p ower of the largest laser line. So, if the largest laser line is
2 dBm and the peak threshold value is 10 dB, the peak threshold limit
is –8 dBm
. You can set the peak threshold
value between 0 to 40 dB.
The peak threshold’s default value is 10 dB. This ensures that any
modulated signals being measured are not confused with their AM
sidebands. For unmodulated lasers, or Fabry- Perot lasers, it may be
desirable to increase this threshold to look for resp onses that are more
than 10 dB from the p eak.
Peak threshold can be used to sup p ress sp urious signals. For examp le,
a laser that is amplitude modulated in the audio frequency range can
cause spurious wavelengths to be displayed below and above the cor-
rect wavelength. The p ower of these sp urious wavelengths is below
that of the correct wavelength. These sp urious signals can be elimi-
nated by decreasing Peak threshold from its Preset value.
8 dBm
–
2 dBm
10 dB
–
=
(
)
Содержание 86120C
Страница 12: ......
Страница 26: ...1 14 Getting Started Returning the Instrument for Service ...
Страница 27: ...1 15 Getting Started Returning the Instrument for Service ...
Страница 28: ......
Страница 96: ...3 18 Programming Monitoring the Instrument ...
Страница 128: ...3 50 Programming Lists of Commands ...
Страница 158: ...4 30 Programming Commands CALCulate1 Subsystem Query Response For normal update 15 047 For fast update 7 525 ...
Страница 236: ...4 108 Programming Commands UNIT Subsystem ...
Страница 248: ......
Страница 262: ...6 14 Specifications and Regulatory Information Declaration of Conformity Declaration of Conformity ...
Страница 264: ...6 16 Specifications and Regulatory Information Product Overview ...
Страница 269: ...7 5 Reference Menu Maps Appl s Menu ...
Страница 271: ...7 7 Reference Menu Maps Display List by WL Menu Delta On Menu ...
Страница 273: ...7 9 Reference Menu Maps System Print Menu ...
Страница 274: ...7 10 Reference Menu Maps System Setup Menu ...
Страница 284: ......
Страница 292: ......
Страница 293: ......