Step 7. Turn Off Wavelength Limiting
Step 7. Turn Off Wavelength Limiting
The instrument’s
Preset
key sets the entire Agilent 86120C wavelength
range of 1270–1650 nm. If a user- defined wavelength range limit was
set using
WL LIM
, the following procedure will ensure that responses
across the full wavelength are measured by returning the instrument to
its preset state.
1
Press the
Setup
key.
2
Press the
WL LIM
softkey.
3
Press
LIM OFF
to remove the limits on wavelength range.
Содержание 86120C
Страница 12: ......
Страница 26: ...1 14 Getting Started Returning the Instrument for Service ...
Страница 27: ...1 15 Getting Started Returning the Instrument for Service ...
Страница 28: ......
Страница 96: ...3 18 Programming Monitoring the Instrument ...
Страница 128: ...3 50 Programming Lists of Commands ...
Страница 158: ...4 30 Programming Commands CALCulate1 Subsystem Query Response For normal update 15 047 For fast update 7 525 ...
Страница 236: ...4 108 Programming Commands UNIT Subsystem ...
Страница 248: ......
Страница 262: ...6 14 Specifications and Regulatory Information Declaration of Conformity Declaration of Conformity ...
Страница 264: ...6 16 Specifications and Regulatory Information Product Overview ...
Страница 269: ...7 5 Reference Menu Maps Appl s Menu ...
Страница 271: ...7 7 Reference Menu Maps Display List by WL Menu Delta On Menu ...
Страница 273: ...7 9 Reference Menu Maps System Print Menu ...
Страница 274: ...7 10 Reference Menu Maps System Setup Menu ...
Страница 284: ......
Страница 292: ......
Страница 293: ......