4-47
Programming Commands
CALCulate3 Subsystem
ASNR:COUNt
Sets the number of measurements to be used for the average signal- to-
noise calculation.
Syntax
:CALCulate3:ASNR:COUNt {?|{<integer> | MINimum | MAXimum }}
<
integer
> is a value that is within the following limits:
Constant
Description
MINimum 10
MAXimum 900
Attribute
Summary
Preset
State: 100
*RST State: 100
SCPI Comp liance: instrument sp ecific
Description
This command sets the number of measurements to be used for the
average signal- to- noise calculation. If this count is changed while the
average signal calculation is on, and the new count is less than the
number of measurements already taken, the instrument will go into
single measurement mode.
Содержание 86120C
Страница 12: ......
Страница 26: ...1 14 Getting Started Returning the Instrument for Service ...
Страница 27: ...1 15 Getting Started Returning the Instrument for Service ...
Страница 28: ......
Страница 96: ...3 18 Programming Monitoring the Instrument ...
Страница 128: ...3 50 Programming Lists of Commands ...
Страница 158: ...4 30 Programming Commands CALCulate1 Subsystem Query Response For normal update 15 047 For fast update 7 525 ...
Страница 236: ...4 108 Programming Commands UNIT Subsystem ...
Страница 248: ......
Страница 262: ...6 14 Specifications and Regulatory Information Declaration of Conformity Declaration of Conformity ...
Страница 264: ...6 16 Specifications and Regulatory Information Product Overview ...
Страница 269: ...7 5 Reference Menu Maps Appl s Menu ...
Страница 271: ...7 7 Reference Menu Maps Display List by WL Menu Delta On Menu ...
Страница 273: ...7 9 Reference Menu Maps System Print Menu ...
Страница 274: ...7 10 Reference Menu Maps System Setup Menu ...
Страница 284: ......
Страница 292: ......
Страница 293: ......