![Agilent Technologies 86120C Скачать руководство пользователя страница 255](http://html.mh-extra.com/html/agilent-technologies/86120c/86120c_user-manual_2867947255.webp)
6-7
Specifications and Regulatory Information
Specifications—NORMAL Update Mode
Input Return Loss
With straight contactconnectors
35 dB
With angled contact connectors (Option 022)
50 dB
Measurement Cycle Time
Normal update mode (characteristic)
1.0 s (1 measurement-per-second)
Measurement Applications
Signal-to-Noise Ratio (characteristic
)
a
channel spacing
≥
100 GHz
>35 dB
channel spacing
≥
50 GHz
>27 dB
Signal-to-Noise Ratio with Averaging, modulated lasers (characteristic)
a
channel spacing
≥
100 GHz
>35 dB with 100 averages
channel spacing
≥
50 GHz
>27 dB with 100 averages
a. 0.1 nm noise bandwidth, lines above –25 dBm.
Содержание 86120C
Страница 12: ......
Страница 26: ...1 14 Getting Started Returning the Instrument for Service ...
Страница 27: ...1 15 Getting Started Returning the Instrument for Service ...
Страница 28: ......
Страница 96: ...3 18 Programming Monitoring the Instrument ...
Страница 128: ...3 50 Programming Lists of Commands ...
Страница 158: ...4 30 Programming Commands CALCulate1 Subsystem Query Response For normal update 15 047 For fast update 7 525 ...
Страница 236: ...4 108 Programming Commands UNIT Subsystem ...
Страница 248: ......
Страница 262: ...6 14 Specifications and Regulatory Information Declaration of Conformity Declaration of Conformity ...
Страница 264: ...6 16 Specifications and Regulatory Information Product Overview ...
Страница 269: ...7 5 Reference Menu Maps Appl s Menu ...
Страница 271: ...7 7 Reference Menu Maps Display List by WL Menu Delta On Menu ...
Страница 273: ...7 9 Reference Menu Maps System Print Menu ...
Страница 274: ...7 10 Reference Menu Maps System Setup Menu ...
Страница 284: ......
Страница 292: ......
Страница 293: ......