Removing a Module
6 Changing the ParBERT Hardware
Agilent 81250 ParBERT Installation Guide, May 2004
6-3
The giga clock connectors do not tolerate any oblique insertion or
removal.
If you need to disconnect one of these cables, use the giga cable removal
tool provided with the system and follow the instructions given below.
Not following these instructions can result in severe damage of the clock
or data module. It can also result in unobtrusive damage of the cable
connector which may then destroy another clock or data module when it
is plugged in.
Grab the cable connector with the tweezers and pull it off, perpendicular
to the module front panel.
Figure 43 Using the Giga Cable Removal Tool
How to Plug Out the Module
To remove the module from the mainframe:
1 Using a 2-mm pozidrive or flathead screwdriver, unscrew the two
screws at the top and bottom of the front panel that hold the module
in place.
2 Flip the module handles until the module moves.
3 Pull the module out of the mainframe.
Содержание 81250
Страница 1: ...Agilent Parallel Bit Error Ratio Tester 81250 ParBERT Installation Guide Agilent Technologies ...
Страница 12: ...1 About this Manual Document History 1 6 Agilent 81250 ParBERT Installation Guide May 2004 ...
Страница 30: ...2 Introduction Unpacking ParBERT 2 18 Agilent 81250 ParBERT Installation Guide May 2004 ...
Страница 138: ...Appendix Declaration of Conformity 10 2 Agilent 81250 ParBERT Installation Guide May 2004 Declaration of Conformity ...
Страница 139: ...Declaration of Conformity Appendix Agilent 81250 ParBERT Installation Guide May 2004 10 3 ...
Страница 141: ...Site Attenuation Requirements Appendix Agilent 81250 ParBERT Installation Guide May 2004 10 5 ...
Страница 142: ...Appendix Site Attenuation Requirements 10 6 Agilent 81250 ParBERT Installation Guide May 2004 ...
Страница 145: ......
Страница 146: ...Copyright Agilent Technologies 2004 Printed in Germany May 2004 81250 91010 Agilent Technologies ...