2 Introduction
ParBERT Components Overview
2-6
Agilent 81250 ParBERT Installation Guide, May 2004
The E4809A clock module needs two VXI slots and provides both clocks
to up to 10 data modules.
Clock to expander frames
Master clock to
data modules
Giga clock to
data modules
External CLOCK INPUT
10 MHz REF.
CLOCK INPUT
TRIGGER OUTPUT
START INPUT
Deskew probe connector
Figure 4 E4809A Clock Module
The E4809A clock module uses power splitters for the giga clock outputs.
The output connectors are paired as follows: [1,2] [3,4] [5,6] [7,8] [9,10]
[11,12]. Output #1 is at the top, output #12 is at the bottom.
If only one connector of a pair is used, the unused connector must be
terminated with 50 Ohm to RF ground.
For example:
•
Only one module is in the system and the corresponding giga clock
cable connected to output #1. Output #2 has to be terminated with
50 Ohm to GND.
•
Two modules are in the system and the corresponding giga clock cables
are connected to output #5 and output #6. No other output needs to be
terminated with 50 Ohm to GND.
NOTE
Содержание 81250
Страница 1: ...Agilent Parallel Bit Error Ratio Tester 81250 ParBERT Installation Guide Agilent Technologies ...
Страница 12: ...1 About this Manual Document History 1 6 Agilent 81250 ParBERT Installation Guide May 2004 ...
Страница 30: ...2 Introduction Unpacking ParBERT 2 18 Agilent 81250 ParBERT Installation Guide May 2004 ...
Страница 138: ...Appendix Declaration of Conformity 10 2 Agilent 81250 ParBERT Installation Guide May 2004 Declaration of Conformity ...
Страница 139: ...Declaration of Conformity Appendix Agilent 81250 ParBERT Installation Guide May 2004 10 3 ...
Страница 141: ...Site Attenuation Requirements Appendix Agilent 81250 ParBERT Installation Guide May 2004 10 5 ...
Страница 142: ...Appendix Site Attenuation Requirements 10 6 Agilent 81250 ParBERT Installation Guide May 2004 ...
Страница 145: ......
Страница 146: ...Copyright Agilent Technologies 2004 Printed in Germany May 2004 81250 91010 Agilent Technologies ...