96
Troubleshooting
8
Troubleshooting Tasks
To Measure FID Leakage Current
1
Load the analytical method.
•
Make sure flows are acceptable for ignition.
•
Heat the detector to operating temperature or to 300 °C.
2
Turn off the FID flame.
3
Verify that the FID electrometer is on.
4
Press
[Front Detector]
or
[Back Detector]
, then scroll to
Output
.
5
Verify that the output is stable and < 1.0 pA.
If the output is unstable or > 1.0 pA, turn off the GC and
check for proper assembly of the upper FID parts and for
contamination. If the contamination is confined to the
detector, bakeout the FID.
6
Turn on the flame.