Troubleshooting
3
Chromatographic Symptoms
FPD Low Peak Areas
•
Check the flow settings, then measure the actual detector
flows. (See “To Measure a Detector Flow” .)
•
Perform complete inlet maintenance: Replace all consumable
parts and bake out the inlet.
•
Perform column maintenance: Bake out contaminants,
remove the contaminated length of column near the inlet,
and reverse and bake out the column as needed.
•
Verify that the column is installed properly.
•
Consider the filter type (sulfur or phosphorus).
•
Check the system for leaks. (See “Checking for Leaks” .)
•
Verify that the method settings are appropriate.
•
Check the flow rates.
•
Check the makeup gas type.