
GC Not Ready Symptoms
4
Troubleshooting
65
FID Does Not Ignite
•
Check that the FID ignitor glows during ignition sequence.
(See To Verify FID Ignitor Function During Ignition
Sequence.)
•
Verify that the Lit Offset is <
2.0 pA.
•
Check for a plugged or partially plugged jet.
•
Check the FID flow rates. The hydrogen:air ratio greatly
impacts ignition. Nonoptimal flow settings can prevent flame
ignition. (See To Measure a Detector Flow.)
•
There could be a large leak in the system if the flame still will
not light. Large leaks result in measured flow rates being
different from actual flow rates, causing nonideal ignition
conditions. Thoroughly leak check the whole system,
especially the column fitting at the FID.
•
Check the column flow rate.
•
Check for leaks at the FID column fitting.
•
Ensure that the FID temperature is high enough for ignition
(>150 °C).
Содержание 6890N
Страница 1: ...Agilent Technologies Agilent 6890N Gas Chromatograph Troubleshooting ...
Страница 20: ...20 Troubleshooting 2 Hardware Symptoms ...
Страница 76: ...76 Troubleshooting 6 Nonfunctional GC Symptoms ...
Страница 88: ...88 Troubleshooting 8 Troubleshooting Tasks ...
Страница 102: ...102 Troubleshooting 8 Troubleshooting Tasks ...