Chromatographic Symptoms
3
Troubleshooting
45
NPD Solvent Quenching
If the baseline does not recover after a solvent peak, try the
following:
•
Turn hydrogren off/on around the solvent peak.
•
Use nitrogen as the makeup gas.
•
Set the total column flow and makeup gas to less than
10 mL/min.
•
Increase the air flow by 10 mL/min.
•
Increase the detector temperature to 325 °C.
•
Implement an Agilent Dean’s switch solvent vent solution.