Chromatographic Symptoms
3
Troubleshooting
39
FID Baseline Output Above 20 pA
•
Verify the purity of the carrier and detector gas supply.
•
Inspect the column for column bleed.
•
Check the gas supply trap indicators/dates and ensure that
the traps are not expended.
•
Verify that the detector was reassembled properly after
recent maintenance.
•
Inspect the detector for contamination.
•
Check that the FID leakage current is < 2.0 pA. (See “To
Measure FID Leakage Current” .)