88 Calibration
Program Listing for "B" Modules
10
ASSIGN @Ld TO 705
20
Chan=l
30
OUTPUT @Ld;”CHAN”;Chan;”;CAL ON"
40
Cal_curr(@Ld,Chan,Hi_curr_rng,Hi_curr_offset,l)
50
Cal_curr(@Ld,Chan,Lo_curr_rng,Lo_curr_offset,0)
60
Cal_volt(@Ld,Chan,Volt_hipt,Volt_lopt)
70
Cal_res(@Ld,Chan,Lo_res_rng,Lo_res_hipt,Lo_res_lopt,0)
80
Cal_res(@Ld,Chan,Mid_res_rng,Mid_res_hipt,Mid_res_lopt,l)
90
Cal_res(@Ld,Chan,Hi_res_rng,Hi_res_hipt,Hi_res_lopt,1)
100
OUTPUT @Ld;"CAL:SAV"
110
OUTPUT @Ld;"CAL OFF"
120
CLEAR SCREEN
130
PRINT TABXY(10,10);"CALIBRATION DONE"
140
END
150
!
160
SUB Cal_curr(@Ld,Chan,Curr_rng,Curr_offset,Flag)
170
PRINT "CURRENT CALIBRATION, RANGE ";Curr_rng
180
PRINT "Set power supply according to calibration information table"
190
PRINT "Use the correct current shunt for the range you are calibrating"
200
PRINT "Press CONT when ready"
210
PAUSE
220
OUTPUT @Ld;"CHAN";Chan
230
OUTPUT @Ld;"MODE:CURR"
240
OUTPUT @Ld;"CURR:RANG";Curr_rng
250
OUTPUT @Ld;"CURR";.05*Curr_rng
260
INPUT "Enter current through shunt for low point in amps",Lopt_curr
270
OUTPUT @Ld;"CAL:LEV:LOW";Lopt_curr
280
OUTPUT @Ld;"CURR";.85*Curr_rng
290
IF Flag THEN WAIT 25
300
INPUT "Enter current through shunt for high point in amps",Hipt_curr
310
OUTPUT @Ld;"CAL:LEV:HIGH";Hipt_curr
320
OUTPUT @Ld;"CURR";Curr_rng
330
INPUT "Enter current through shunt for high point in amps",Hipt_curr
340
OUTPUT @Ld;"CAL:LEV:HIGH";Hipt_curr
350
IF Flag THEN OUTPUT @Ld;"CAL:MEAS:HIGH";Hipt_curr
360
IF Flag THEN
370
OUTPUT @Ld;"CURR";4*(Curr_rng/3750)
380
WAIT 25
390
INPUT "Enter current through shunt for low point in amps",Lopt_curr
400
OUTPUT @Ld;"CAL:LEV:LOW";(Lopt_curr-Curr_offset)
410
OUTPUT @Ld;"CAL:MEAS:LOW";Lopt_curr
420
ELSE
430
OUTPUT @Ld;"CURR";10*(Curr_rng/3750)
440
INPUT "Enter current through shunt for low point in amps",Lopt_curr
450
OUTPUT @Ld;"CAL:LEV:LOW";(Lopt_curr-Curr_offset)
460
END IF
470
PRINT "Test unit to verify that program and readback values are in spec"
480
PRINT "Press CONT when ready to calibrate transient levels
490
PAUSE
500
OUTPUT @Ld;"CURR";.05*Curr_rng
510
OUTPUT @Ld;"TRAN:STAT ON;MODE TOGG;:TRIG:SOUR BUS"
520
OUTPUT @Ld;"CURR:TLEV";.85*Curr_rng
Содержание 6050A
Страница 10: ......
Страница 14: ......
Страница 32: ......
Страница 46: ...46 Installation Figure 3 12 Local Sensing Figure 3 13 Remote Sensing...
Страница 47: ...Installation 47 Figure 3 14 Parallel Operation Figure 3 15 Zero Volt Loading...
Страница 48: ......
Страница 55: ...Local Operation 55 Figure 4 2 Recommended Programming Sequence...
Страница 64: ......
Страница 69: ...Remote Operation 69 Figure 5 1 Remote Programming Flowchart Sheet 1...
Страница 70: ...70 Remote Operation Figure 5 1 Remote Programming Flowchart Sheet 2...
Страница 74: ......
Страница 78: ...78 Calibration Figure 6 2 Calibration Flowchart for A Modules...
Страница 79: ...Calibration 79 Figure 6 2 Calibration Flowchart for A Modules continued...
Страница 80: ...80 Calibration Figure 6 2 Calibration Flowchart for A Modules continued...
Страница 85: ...Calibration 85 Figure 6 3 Calibration Flowchart for B Modules...
Страница 86: ...86 Calibration Figure 6 3 Calibration Flowchart for B Modules continued...
Страница 87: ...Calibration 87 Figure 6 3 Calibration Flowchart for B Modules continued...