96
34405A User’s and Service Guide
4
Performance Tests and Calibration
6
Enter the actual applied input (see
“Entering Adjustment Values”
on
page 86).
7
Press
to start the adjustment. The
CAL
annunciator flashes to
indicate the calibration is in progress.
•
Successful completion of each adjustment value is indicated by a
short beep and the primary display briefly showing
PASS
.
•
An adjustment failure is indicated by a long beep, the primary
display showing
FAiL
and a calibration error number appearing in
the secondary display. Check the input value, range, function, and
entered adjustment value to correct the problem and repeat the
adjustment step.
8
Repeat steps 3 through 7 for each gain adjustment point shown in the
table.
9
Verify the AC Current Gain adjustments using the
“AC Current
Verification Test”
on page 78.
Table 18
AC Current Gain Adjustment
Input
Current
Frequency
Function
Frequency as 1kHz
Cal Item
1 mA
1kHz
AC Current
1 mA
10 mA
1kHz
10 mA
100 mA
1kHZ
100 mA
1000 mA
1kHz
1000 mA
Caution: Connect calibrator to multimeter’s 12A and LO
terminals before applying the following 1A and 10A
1 A
1kHz
1 A
10 A
1kHz
10 A
Disp
Downloaded from
www.Manualslib.com
manuals search engine
Содержание 34405A
Страница 122: ...122 34405A User s and Service Guide 5 Disassembly and Repair Downloaded from www Manualslib com manuals search engine...
Страница 138: ...138 34405A User s and Service Guide 6 Specifications Downloaded from www Manualslib com manuals search engine...
Страница 142: ...142 34405A User s and Service Guide Index Downloaded from www Manualslib com manuals search engine...