34405A User’s and Service Guide
109
Disassembly and Repair
5
To Replace the Power Line Fuse
The power line fuse is located within the multimeter’s fuse–holder
assembly on the rear panel. The multimeter is shipped from the factory
with a power–line fuse installed. The supplied fuse is a time- lag,
low- breaking, 0.2A/ 250V, 5x20mm fuse, Agilent part number
2110- 1395. If
you determine that the fuse is faulty, replace it with one of the same size
and rating.
2
1
1
Disconnect power cord. Depress tabs 1 and
2 and pull fuse holder from rear panel.
2
Remove the line voltage selector from fuse
holder assembly.
3
Rotate line voltage selector and reinstall so
correct voltage appears in fuse holder
window.
4
Replace fuse holder assembly in rear
panel.
100, 120, 220 (230) or 240 Vac