82
34405A User’s and Service Guide
4
Performance Tests and Calibration
Calibration Security
The calibration security code prevents accidental or unauthorized
adjustments to the instrument. When you first receive your instrument, it
is secured. Before you can adjust the instrument, you must unsecure it by
entering the correct security code (see
“Unsecuring the Instrument for
Calibration”
on page 83).
The security code is set to AT34405 when the instrument is shipped from
the factory. The security code is stored in non- volatile memory, and does
not change when power is turned off or after a Factory Reset (*RST
command) or after an Instrument Preset (SYSTem:PRESet command).
The security code may contain up to 12 alphanumeric characters. The first
character must be a letter. The remaining characters can be letters or
numbers. You do not have to use all 12 characters.
N O T E
You can unsecure the instrument from the front panel, but you cannot enter a new security
code or change the security code from the front panel. The security code can only be
changed from the remote interface after the instrument has been unsecured. Refer to the
CAL:SEC:CODE command in the 34405A Programmer's Reference Help File for more details.
Downloaded from
www.Manualslib.com
manuals search engine
Содержание 34405A
Страница 122: ...122 34405A User s and Service Guide 5 Disassembly and Repair Downloaded from www Manualslib com manuals search engine...
Страница 138: ...138 34405A User s and Service Guide 6 Specifications Downloaded from www Manualslib com manuals search engine...
Страница 142: ...142 34405A User s and Service Guide Index Downloaded from www Manualslib com manuals search engine...