80
34405A User’s and Service Guide
4
Performance Tests and Calibration
Optional AC Current Performance Verification Test
Configuration: AC Current (CONFigure:CURRent:AC)
1
Select the AC Current function.
2
Select each range in the order shown below. Provide the indicated input
voltage and frequency. Compare measurement results to the appropriate
test limits shown in the table. (Be certain to allow for appropriate
source settling.)
[1] Select 5½- digit measurement resolution
Table 12
Optional AC Current Performance Verification Test
Current
Input Frequency Function
[1]
Range
Error from Nominal
1 year
10mA
20Hz
AC Current
10mA
± 0.16mA
10mA
45Hz
10mA
± 0.16mA
10mA
1kHz
10mA
± 60µA
10mA
10kHz
10mA
± 0.22mA
1A
1kHz
1A
± 6mA
100mA
1kHz
1A
± 1.5mA
10mA
1kHz
1A
± 1.05mA