Performance Tests and Calibration
4
34405A User’s and Service Guide
71
Zero Offset Verification
This test is used to check the zero offset performance of the instrument.
Verification checks are only performed for those functions and ranges with
unique offset calibration constants. Measurements are checked for each
function and range as described in the procedure on the next page.
Zero Offset Verification Test
1
Connect the Shorting Plug to the HI and LO input terminals. (see
“Input
Connections”
on page 67). Leave the current inputs open.
2
Select each function and range in the order shown in the table below.
Make a measurement and observe the result. Compare measurement
results to the appropriate test limits shown in the table below (table
continued on thefollowing page).
N O T E
Note that resistance measurements use the Null math function (Null reading taken with
test leads connected together) to eliminate test lead resistance.
Table 4
Zero Offset Verification Test
Input
Function
[1]
Range
Quick Check
Error from Nominal
1 year
Open
DC Current
10mA
Q
±1.5µA
Open
100mA
±5 µA
Open
1A
±70µA
Open
10A
±0.7mA
Open
Capacitance
1nF
±8pF
Open
10nF
±0.05nF
Open
100nF
±0.5nF
Open
1µF
±5nF
Open
10µF
±0.05µF
Open
100µF
±0.5µF
Downloaded from
www.Manualslib.com
manuals search engine
Содержание 34405A
Страница 122: ...122 34405A User s and Service Guide 5 Disassembly and Repair Downloaded from www Manualslib com manuals search engine...
Страница 138: ...138 34405A User s and Service Guide 6 Specifications Downloaded from www Manualslib com manuals search engine...
Страница 142: ...142 34405A User s and Service Guide Index Downloaded from www Manualslib com manuals search engine...