![Agilent Technologies 2200 TapeStation Скачать руководство пользователя страница 62](http://html1.mh-extra.com/html/agilent-technologies/2200-tapestation/2200-tapestation_user-manual_2867581062.webp)
62
2200 TapeStation
6
Troubleshooting
Troubleshooting the Test Kit Manufacturers TestKitSuite.xml File
Fragments can be
seen in the gel image
(marked with a red
arrow in this
example) that do not
have a pathogen
marker sign next to
them and are not
reported in the result
table.
The
TestKitSuite.xml
file, provided by the PCR
Test Kit Manufacturer, has the ability to ignore
expected fragments. This then ensures that they
are not presented when that fragment is detected.
This is usually used for known non-specific
amplification and/or primer dimers.
Run the sample
again. If the issue
re-occurs, contact
the Test Kit
Manufacturer.
Table 9
Overview troubleshooting test kit manufacturers TestKitSuite.xml file
Image
Issue
Probable cause
Solution