Maintenance
5
Changing the Needle
Change the needle cartridge
1
Remove the sample plate and tip holder.
2
Remove the foil tab from the top of the needle cartridge.
3
Insert the needle cartridge into the tip holder space, using the label for
orientation. The cartridge should be placed so that the label faces to the
right, and the printed arrow points to the front of the TapeStation.
4
Close the lid.
5
Go to
Needle Change
on the Controller software toolbar and select
Run
.
Parts required
#
p/n
Description
1
G2960-60062
Needle cartridge (for use in single pump systems)
For use with product numbers ST007, ST008 and ST009
OR
1
G2960-60063
Needle cartridge (for use in dual pump systems)
For use with product numbers ST017, ST019, ST010, G2960A,
G2961A, G2964AA, G2965AA and G2966AA
N O T E
New needles cartridges can be ordered at any time from Agilent Technologies by
contacting your local sales agent.
For details on correct needle cartridge for your TapeStation model, refer to
N O T E
Care must be taken to keep the needle cartridge level after removing the foil tab
Needle cartridge
Front
Rear
TapeStation
Содержание 2200 TapeStation
Страница 1: ...Agilent Technologies Agilent 2200 TapeStation User Manual...
Страница 10: ...10 2200 TapeStation 1 Introduction to the 2200 TapeStation System Overview of the System...
Страница 14: ...14 2200 TapeStation 2 Specifications Technical Specifications...
Страница 42: ...42 2200 TapeStation 4 Using the 2200 TapeStation System DNA Sample Preparation...
Страница 48: ...48 2200 TapeStation 5 Maintenance Changing the Needle...
Страница 64: ...64 2200 TapeStation 6 Troubleshooting Troubleshooting the Test Kit Manufacturers TestKitSuite xml File...
Страница 71: ......