![Agilent Technologies 2200 TapeStation Скачать руководство пользователя страница 55](http://html1.mh-extra.com/html/agilent-technologies/2200-tapestation/2200-tapestation_user-manual_2867581055.webp)
2200 TapeStation
55
Troubleshooting
6
ScreenPlex Software
ScreenPlex Software
Table 7
ScreenPlex Software
Issue
Probable cause
Solution
Upon result launch, a
message states that
the software failed to
map a suitable top
and bottom marker.
There may be an
error in the ladder
lane – the bands are
too faint or the lane
is blank.
Perform the entire run, including samples, on a
new piece of ScreenTape.
Upon result launch, a
message states that
the software failed to
map detected peaks
with expected
controls.
There may be an
error in the ladder
lane – the ladder has
too many or too few
bands than expected.
Perform the entire run, including samples, on a
new piece of ScreenTape.
Unidentified peaks
have been detected
in the sample.
There may be the
presence of an
unknown fragment in
the sample or a
particle has adhered
to the outside of
ScreenTape.
Freshly prepare the sample again and re-run on a
new piece of ScreenTape.
Re-running a sample
that has been left in
the 2200 TapeStation
has provided a
different result to the
previous run.
The samples were
not prepared
optimally.
Samples should be mixed with loading buffer,
then vortexed for 5 seconds and spun down
before placing in the 2200 TapeStation.
A completely blank
lane has appeared in
the results.
There may be a
bubble in the sample.
Briefly centrifuge the sample to remove any
bubbles and re-run on a new piece of ScreenTape.
If the problem persists, please contact Agilent
Technologies.
All markers and
sample bands within
a lane appear very
faint.
There may be a
bubble in the sample.
Briefly centrifuge the sample to remove any
bubbles and re-run on a new piece of ScreenTape.
If the problem persists, please contact Agilent
Technologies.
Содержание 2200 TapeStation
Страница 1: ...Agilent Technologies Agilent 2200 TapeStation User Manual...
Страница 10: ...10 2200 TapeStation 1 Introduction to the 2200 TapeStation System Overview of the System...
Страница 14: ...14 2200 TapeStation 2 Specifications Technical Specifications...
Страница 42: ...42 2200 TapeStation 4 Using the 2200 TapeStation System DNA Sample Preparation...
Страница 48: ...48 2200 TapeStation 5 Maintenance Changing the Needle...
Страница 64: ...64 2200 TapeStation 6 Troubleshooting Troubleshooting the Test Kit Manufacturers TestKitSuite xml File...
Страница 71: ......