36
2200 TapeStation
4
Using the 2200 TapeStation System
Operating Procedure
How to Use the ScreenPlex Analysis Software
Once the imaging is complete, the analysis software will automatically launch
the ScreenPlex Result File (extension .srf). The data provided includes the test
types selected (displayed in tabs), the date and time at which the samples
were run, the tape serial number and the system information. In addition,
each sample result and sample ID will appear.
•
If a band is detected a
+
sign in a grey box is displayed, if no band is
detected a
-
sign in a white box will be displayed.
•
The numerical values next to the
Detected Bands
represent intensity levels
relative to the top marker, which has arbitrarily been assigned a value of
100 .
•
Any absent internal controls and/or unidentified bands are detailed in the
table and highlighted with a warning sign. The gel image also contains the
corresponding warning signs. The image below illustrates both cases.
•
The analysed gel image can be viewed by selecting the
View Gel
tick box in
the top right hand corner of the
ScreenPlex Result Viewer
. The gel image
shows the top and bottom marker selection (using a < symbol) as well as the
sample band detection (using a grey arrow). Any unidentified bands are
Содержание 2200 TapeStation
Страница 1: ...Agilent Technologies Agilent 2200 TapeStation User Manual...
Страница 10: ...10 2200 TapeStation 1 Introduction to the 2200 TapeStation System Overview of the System...
Страница 14: ...14 2200 TapeStation 2 Specifications Technical Specifications...
Страница 42: ...42 2200 TapeStation 4 Using the 2200 TapeStation System DNA Sample Preparation...
Страница 48: ...48 2200 TapeStation 5 Maintenance Changing the Needle...
Страница 64: ...64 2200 TapeStation 6 Troubleshooting Troubleshooting the Test Kit Manufacturers TestKitSuite xml File...
Страница 71: ......