Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
2
Horizontal Controls
Negative delay values indicate you're looking at a portion of the waveform
before the trigger event, and positive values indicate you're looking at the
waveform after the trigger event.
To change the time/div of the normal window, turn off Zoom; then, turn
the horizontal scale (sweep speed) knob.
For information about using zoom mode for measurements, refer to
"To
isolate a pulse for Top measurement"
on page 175 and
"To isolate an event
for frequency measurement"
on page 182.
To change the horizontal scale knob's coarse/fine adjustment setting
1
Push the horizontal scale knob (or press
[Horiz] > Fine
) to toggle between
fine and coarse adjustment of the horizontal scale.
When
Fine
is enabled, turning the horizontal scale knob changes the
time/div (displayed in the status line at the top of the display) in smaller
increments. The time/div remains fully calibrated when
Fine
is on.
When
Fine
is turned off, the Horizontal scale knob changes the time/div
setting in a 1- 2- 5 step sequence.
To position the time reference (left, center, right)
Time reference is the reference point on the display for delay time
(horizontal position).
1
Press
[Horiz]
.
2
In the Horizontal Menu, press
Time Ref
; then, select:
•
Left
— the time reference is set to one major division from the left
edge of the display.
•
Center
— the time reference is set to the center of the display.
•
Right
— the time reference is set to one major division from the right
edge of the display.
A small hollow triangle (
∇
) at the top of the display grid marks the
position of the time reference. When delay time is set to zero, the trigger
point indicator (
▼
) overlays the time reference indicator.
Содержание 2000 X
Страница 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Страница 18: ...18 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Страница 42: ...42 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Страница 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Страница 76: ...76 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Страница 100: ...100 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Страница 106: ...106 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Display Settings ...
Страница 134: ...134 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 9 Triggers ...
Страница 142: ...142 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Trigger Mode Coupling ...
Страница 168: ...168 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 12 Cursors ...
Страница 204: ...204 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Digital Voltmeter ...
Страница 216: ...216 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Страница 230: ...230 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Save Recall Setups Screens Data ...
Страница 236: ...236 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Print Screens ...
Страница 272: ...272 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Web Interface ...
Страница 290: ...290 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Reference ...