11
Acquisition Control
Oscilloscope Bandwidth Required
The oscilloscope bandwidth required to accurately measure a signal is
primarily determined by the signal's rise time, not the signal's frequency.
You can use these steps to calculate the oscilloscope bandwidth required:
1
Determine the fastest edge speeds.
You can usually obtain rise time information from published
specifications for devices used in your designs.
2
Compute the maximum "practical" frequency component.
From Dr. Howard W. Johnson's book,
High- Speed Digital Design – A
Handbook of Black Magic
, all fast edges have an infinite spectrum of
frequency components. However, there is an inflection (or "knee") in the
frequency spectrum of fast edges where frequency components higher
than f
knee
are insignificant in determining the shape of the signal.
f
knee
= 0.5 / signal rise time (based on 10% - 90% thresholds)
f
knee
= 0.4 / signal rise time (based on 20% - 80% thresholds)
3
Use a multiplication factor for the required accuracy to determine the
oscilloscope bandwidth required.
See Also
Choosing an Oscilloscope with the Right Bandwidth for your
Application
, Agilent Application Note 1588
(
"http://cp.literature.agilent.com/litweb/pdf/5989- 5733EN.pdf"
)
Required accuracy
Oscilloscope bandwidth required
20%
f
BW
= 1.0 x f
knee
10%
f
BW
= 1.3 x f
knee
3%
f
BW
= 1.9 x f
knee
Содержание 2000 X
Страница 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Страница 18: ...18 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Страница 42: ...42 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Страница 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Страница 76: ...76 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Страница 100: ...100 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Страница 106: ...106 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Display Settings ...
Страница 134: ...134 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 9 Triggers ...
Страница 142: ...142 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Trigger Mode Coupling ...
Страница 168: ...168 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 12 Cursors ...
Страница 204: ...204 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Digital Voltmeter ...
Страница 216: ...216 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Страница 230: ...230 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Save Recall Setups Screens Data ...
Страница 236: ...236 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Print Screens ...
Страница 272: ...272 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Web Interface ...
Страница 290: ...290 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Reference ...