Acquisition Control
11
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
147
So, in practice, an oscilloscope's sample rate should be four or more times
its bandwidth: f
S
= 4f
BW
. This way, there is less aliasing, and aliased
frequency components have a greater amount of attenuation.
See Also
Evaluating Oscilloscope Sample Rates vs. Sampling Fidelity: How to
Make the Most Accurate Digital Measurements
, Agilent Application Note
1587 (
"http://cp.literature.agilent.com/litweb/pdf/5989- 5732EN.pdf"
)
Oscilloscope Rise Time
Closely related to an oscilloscope's bandwidth specification is its rise time
specification. Oscilloscopes with a Gaussian- type frequency response have
an approximate rise time of 0.35/f
BW
based on a 10% to 90% criterion.
An oscilloscope's rise time is not the fastest edge speed that the
oscilloscope can accurately measure. It is the fastest edge speed the
oscilloscope can possibly produce.
Figure 23
Sample Rate and Oscilloscope Bandwidth
Limiting oscilloscope bandwidth (f
BW
) to 1/4 the sample rate (f
S
/4)
reduces frequency components above the Nyquist frequency (f
N
).
I
6
I
1
I
6
G%
$WWHQXDWLRQ
$OLDVHGIUHTXHQF\
FRPSRQHQWV
)UHTXHQF\
G%
Содержание 2000 X
Страница 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Страница 18: ...18 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Страница 42: ...42 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Страница 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Страница 76: ...76 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Страница 100: ...100 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Страница 106: ...106 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Display Settings ...
Страница 134: ...134 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 9 Triggers ...
Страница 142: ...142 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Trigger Mode Coupling ...
Страница 168: ...168 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 12 Cursors ...
Страница 204: ...204 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Digital Voltmeter ...
Страница 216: ...216 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Страница 230: ...230 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Save Recall Setups Screens Data ...
Страница 236: ...236 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Print Screens ...
Страница 272: ...272 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Web Interface ...
Страница 290: ...290 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Reference ...