2
Check the data pulse width. Using the oscilloscope, verify that the data pulse width
is 4.000 ns,
+0 ps or
−
100 ps.
a
In the oscilloscope Timebase menu, select Sweep Speed 1.00 ns/div.
b
Select Delay. Using the oscilloscope knob, position the data waveform (Channel 1) so
that the waveform is centered on the screen.
c
In the oscilloscope Delta V menu, set the Marker 1 Position to Chan 1, then set Marker
1 at
−
1.3000 V. Set the Marker 2 Position to Chan 1, then set Marker 2 at
−
1.3000 V.
d
In the oscilloscope Delta T menu, select Start On Pos Edge 1. Select Stop on Neg
Edge 1. Select Precision Edge Find.
e
If the pulse width is outside of the limits, adjust the pulse generator channel 1 width
and select the oscilloscope Precision Edge Find until the pulse width is within limits.
Testing Performance
To test the single-clock, multiple-edge, state acquisition
3–50
Содержание 1664A
Страница 3: ...The Agilent Technologies 1664A Logic Analyzer iii...
Страница 8: ...Contents viii...
Страница 22: ...2 6...
Страница 91: ...Troubleshooting Flowchart 1 Troubleshooting To use the flowcharts 5 3...
Страница 92: ...Troubleshooting Flowchart 2 Troubleshooting To use the flowcharts 5 4...
Страница 93: ...Troubleshooting Flowchart 3 Troubleshooting To use the flowcharts 5 5...
Страница 94: ...Troubleshooting Flowchart 4 Troubleshooting To use the flowcharts 5 6...
Страница 95: ...Troubleshooting Flowchart 5 Troubleshooting To use the flowcharts 5 7...
Страница 96: ...Troubleshooting Flowchart 6 Troubleshooting To use the flowcharts 5 8...
Страница 97: ...Troubleshooting Flowchart 7 Troubleshooting To use the flowcharts 5 9...
Страница 98: ...Troubleshooting Flowchart 8 Troubleshooting To use the flowcharts 5 10...
Страница 99: ...Troubleshooting Flowchart 9 Troubleshooting To use the flowcharts 5 11...
Страница 100: ...Troubleshooting Flowchart 10 Troubleshooting To use the flowcharts 5 12...
Страница 101: ...Troubleshooting Flowchart 11 Troubleshooting To use the flowcharts 5 13...
Страница 102: ...Troubleshooting Flowchart 12 Troubleshooting To use the flowcharts 5 14...
Страница 124: ...Exploded View of the 1664A Replacing Assemblies 6 4...
Страница 137: ...Exploded View Exploded view of the 1664A logic analyzer Replaceable Parts Exploded View 7 3...
Страница 160: ...8 18...
Страница 162: ......