4
Enter the Sys PV tests, then enter the Display Test.
A grid pattern should appear.
5
If the display is tilted (rotated), adjust the CRT yoke by rotating it to straighten the
display.
6
If the grid pattern is not centered horizontally, adjust the H-Hold.
7
If you need to adjust the intensity, go to the next page.
If you are finished with the adjustments, turn off the instrument, then remove the
power cord. Install the cover on the instrument.
Yoke
Calibrating and Adjusting
To adjust the CRT monitor alignment
4–4
Содержание 1664A
Страница 3: ...The Agilent Technologies 1664A Logic Analyzer iii...
Страница 8: ...Contents viii...
Страница 22: ...2 6...
Страница 91: ...Troubleshooting Flowchart 1 Troubleshooting To use the flowcharts 5 3...
Страница 92: ...Troubleshooting Flowchart 2 Troubleshooting To use the flowcharts 5 4...
Страница 93: ...Troubleshooting Flowchart 3 Troubleshooting To use the flowcharts 5 5...
Страница 94: ...Troubleshooting Flowchart 4 Troubleshooting To use the flowcharts 5 6...
Страница 95: ...Troubleshooting Flowchart 5 Troubleshooting To use the flowcharts 5 7...
Страница 96: ...Troubleshooting Flowchart 6 Troubleshooting To use the flowcharts 5 8...
Страница 97: ...Troubleshooting Flowchart 7 Troubleshooting To use the flowcharts 5 9...
Страница 98: ...Troubleshooting Flowchart 8 Troubleshooting To use the flowcharts 5 10...
Страница 99: ...Troubleshooting Flowchart 9 Troubleshooting To use the flowcharts 5 11...
Страница 100: ...Troubleshooting Flowchart 10 Troubleshooting To use the flowcharts 5 12...
Страница 101: ...Troubleshooting Flowchart 11 Troubleshooting To use the flowcharts 5 13...
Страница 102: ...Troubleshooting Flowchart 12 Troubleshooting To use the flowcharts 5 14...
Страница 124: ...Exploded View of the 1664A Replacing Assemblies 6 4...
Страница 137: ...Exploded View Exploded view of the 1664A logic analyzer Replaceable Parts Exploded View 7 3...
Страница 160: ...8 18...
Страница 162: ......