![Agilent Technologies 1168/9A Скачать руководство пользователя страница 79](http://html.mh-extra.com/html/agilent-technologies/1168-9a/1168-9a_user-manual_2867526079.webp)
Using Probe Heads
2
Soldering a ZIF Tip to a DUT
1168/9A Probes User’s Guide
79
Soldering a ZIF Tip to a DUT
Overview
Soldering a ZIF tip into a DUT is straightforward, but some
of the traditional soldering techniques that are typically used
on larger components will not work well here. The following
is an overview of the recommended soldering techniques
1
Add some solder to the DUT connection points. There should
be enough solder to provide a good fillet around the ZIF
tip’s leads, but not so much as to create a big solder ball. A
fine MetCal (or equivalent) soldering tip should be used
along with some 11 or 15 mil solder.
2
Using a rosin flux pen, coat the solder points with flux. The
flux core solder does not provide enough flux for this small
scale soldering. Also, put flux on the tips of the leads of the
ZIF tip.
3
Clean the soldering tip well, then add a little bit of solder to
the tip. It may take several tries to get just a little bit of
solder right at or near the tip of the soldering iron. The
solder on the tip keeps the soldering iron tip from pulling
solder off the DUT connection points. This step may be
optional if there is already enough solder on the DUT
connection points.
4
Position a lead of the ZIF tip on top of one of the target
points, then briefly touch the soldering iron tip to the joint.
The thermal mass of this joint is very small, so you don't
need to dwell on the joint for very long. The flux that was
added to the joint should produce a good, clean solder joint.
If you do not get a good, shinny, strong solder joint, then
there was either not enough flux or the joint was heated too
long and the flux boiled off.
5
There is a possibility that if a lead of the ZIF tip is inserted
into a large ball of solder that is heated excessively with a
soldering iron, the solder joint holding the lead onto the ZIF
tip pc board could flow and the lead would come off
destroying the ZIF tip. Only the first third of the lead or so
needs to be soldered to the target point.
Содержание 1168/9A
Страница 1: ...Agilent Technologies Agilent 1168 9A Differential and Single Ended Probes User s Guide ...
Страница 5: ...Contents 1168 9A Probes User s Guide 5 N5381A and N5382A Probe Heads 248 Other Accessories 249 ...
Страница 6: ...6 1168 9A Probes User s Guide Contents ...
Страница 9: ...Getting Started 1 Introduction 1168 9A Probe s User s Guide 9 Figure 2 Available Probe Heads ...
Страница 40: ...40 1168 9A Probe s User s Guide 1 Getting Started Service ...
Страница 69: ...Using Probe Heads 2 Recommended Configurations 1168 9A Probes User s Guide 69 Figure 36 Removing the Probe ...
Страница 104: ...104 1168 9A Probes User s Guide 2 Using Probe Heads N2887A 8A Soft Touch Probe Heads ...
Страница 120: ...120 1168 9A Probes User s Guide 3 Maintaining Probe Heads E2677A 9A Solder In Probe Heads ...
Страница 136: ...136 1168 9A Probes User s Guide 4 Calibrating Probes N2887A N2888A Calibration and Deskew Procedure ...
Страница 150: ...150 1168A 9A Probe s User s Guide 5 Characteristics and Specifications Probe Dimensions ...
Страница 202: ...202 1168A 9A Probe s User s Guide 6 Performance Plots N2887A N2888A Soft Touch Probe Heads ...
Страница 250: ...250 1168A 9A Probe s User s Guide 9 Replacement Parts Other Accessories ...