1168/9A Probe’s User’s Guide
1
Getting Started
Probe Handling
Connecting the Probe to an
Oscilloscope
The probes are only meant to be plugged into gold plated
BNCs (like those on Infiniium oscilloscopes). To connect the
probe to the oscilloscope, do the following steps:
1
As shown in
Figure 11
, with the lever in the relaxed
position
push the probe onto the BNC. The lever moves
towards the
R
(release)
and returns to the
symbol.
2
Move the lever towards the
symbol until snug.
NOTE
How far the locking mechanism can be pushed to the right varies and will
not be the same for every user. Therefore, do not try to force it further to
the right because you believe it is unlocked. Instead, gently push it until it
is snug.
To disconnect the probe, move and hold the lever at
R
(release)
and pull the probe from the BNC.
Figure 11
Properly Connecting a Probe to the Oscilloscope
Securing Probe Heads and
Amplifiers to Your DUTs
When soldering a probe head to a circuit, first provide strain
relief by using low temperature hot glue (use as little as
possible) or non- conductive double- sided tape. Do not use
super glue and do not get the low temperature hot glue on
the actual probe head tip as this can damage the precision
components of your probing system (only use the low
temperature hot glue on the probe head cables). The
provided velcro pads can be used to secure your probe
amplifier casing to the board.
Once strain relief has been provided, solder the probe tip to
the circuit board and then plug the probe head into the
probe amplifier.
Содержание 1168/9A
Страница 1: ...Agilent Technologies Agilent 1168 9A Differential and Single Ended Probes User s Guide ...
Страница 5: ...Contents 1168 9A Probes User s Guide 5 N5381A and N5382A Probe Heads 248 Other Accessories 249 ...
Страница 6: ...6 1168 9A Probes User s Guide Contents ...
Страница 9: ...Getting Started 1 Introduction 1168 9A Probe s User s Guide 9 Figure 2 Available Probe Heads ...
Страница 40: ...40 1168 9A Probe s User s Guide 1 Getting Started Service ...
Страница 69: ...Using Probe Heads 2 Recommended Configurations 1168 9A Probes User s Guide 69 Figure 36 Removing the Probe ...
Страница 104: ...104 1168 9A Probes User s Guide 2 Using Probe Heads N2887A 8A Soft Touch Probe Heads ...
Страница 120: ...120 1168 9A Probes User s Guide 3 Maintaining Probe Heads E2677A 9A Solder In Probe Heads ...
Страница 136: ...136 1168 9A Probes User s Guide 4 Calibrating Probes N2887A N2888A Calibration and Deskew Procedure ...
Страница 150: ...150 1168A 9A Probe s User s Guide 5 Characteristics and Specifications Probe Dimensions ...
Страница 202: ...202 1168A 9A Probe s User s Guide 6 Performance Plots N2887A N2888A Soft Touch Probe Heads ...
Страница 250: ...250 1168A 9A Probe s User s Guide 9 Replacement Parts Other Accessories ...