Calibrating Probes
4
Calibration for Hand-Held Browser Probe Heads
1168/9A Probes User’s Guide
131
Calibration for Hand-Held Browser Probe Heads
Calibration of the hand- held browser probe heads consists of
a vertical calibration and a skew calibration. The vertical
calibration should be performed before the skew calibration.
Both calibrations should be performed for best probe
measurement performance.
NOTE
Before calibrating the probe, verify that the Infiniium oscilloscope has
been calibrated recently and that the calibration
Δ
temperature is within
±5 °C. If this is not the case, calibrate the oscilloscope before calibrating
the probe. This information is found in Infiniium Calibration dialog box.
Calibration Setup
The calibration procedure requires the following parts.
■
BNC (male) to SMA (male) adaptor
■
Deskew fixture
■
50
Ω
SMA terminator
1
As shown in
on page 132, connect BNC (male) to
SMA (male) adaptor to the deskew fixture on the connector
closest to the yellow pincher.
2
Connect the 50
Ω
SMA terminator to the connector farthest
from the yellow pincher.
3
Connect the BNC side of the deskew fixture to the Aux Out
of the Infiniium oscilloscope.
4
Connect the probe to an oscilloscope channel.
5
Place the positive resistor tip of the browser on the center
conductor of the deskew fixture between the green line and
front end of the yellow pincher. The negative resistor tip or
ground pin of the browser must be on either of the two
outside conductors (ground) of the deskew fixture.
6
On the Infiniium oscilloscope in the Setup menu, select the
channel connected to the probe.
Содержание 1168/9A
Страница 1: ...Agilent Technologies Agilent 1168 9A Differential and Single Ended Probes User s Guide ...
Страница 5: ...Contents 1168 9A Probes User s Guide 5 N5381A and N5382A Probe Heads 248 Other Accessories 249 ...
Страница 6: ...6 1168 9A Probes User s Guide Contents ...
Страница 9: ...Getting Started 1 Introduction 1168 9A Probe s User s Guide 9 Figure 2 Available Probe Heads ...
Страница 40: ...40 1168 9A Probe s User s Guide 1 Getting Started Service ...
Страница 69: ...Using Probe Heads 2 Recommended Configurations 1168 9A Probes User s Guide 69 Figure 36 Removing the Probe ...
Страница 104: ...104 1168 9A Probes User s Guide 2 Using Probe Heads N2887A 8A Soft Touch Probe Heads ...
Страница 120: ...120 1168 9A Probes User s Guide 3 Maintaining Probe Heads E2677A 9A Solder In Probe Heads ...
Страница 136: ...136 1168 9A Probes User s Guide 4 Calibrating Probes N2887A N2888A Calibration and Deskew Procedure ...
Страница 150: ...150 1168A 9A Probe s User s Guide 5 Characteristics and Specifications Probe Dimensions ...
Страница 202: ...202 1168A 9A Probe s User s Guide 6 Performance Plots N2887A N2888A Soft Touch Probe Heads ...
Страница 250: ...250 1168A 9A Probe s User s Guide 9 Replacement Parts Other Accessories ...