![Agilent Technologies 1168/9A Скачать руководство пользователя страница 143](http://html.mh-extra.com/html/agilent-technologies/1168-9a/1168-9a_user-manual_2867526143.webp)
Characteristics and Specifications
5
InfiniiMax II Series with N5380B SMA Probe Head
1168A/9A Probe’s User’s Guide
143
Table 24
Characteristics
Item
1169A
1168A
Bandwidth
>12GHz
> 10GHz
Probe only rise and fall times
27.5 ps (20% to 80%)
40 ps (10% to 90%)
27.5 ps (20% to 80%)
40 ps (10% to 90%)
System rise and fall times
a
With DSO/DSA91304A:
23 ps (20% to 80%)
33 ps (10% to 90%)
With DSO81004A:
30 ps (20% to 80%)
42 ps (10% to 90%)
With DSO/DSA91204A
25 ps (20% to 80%)
36 ps (10% to 90%)
With DSO/DSA90804A:
38 ps (20% to 80%)
54 ps (10% to 90%)
System bandwidth (–3 dB)
With DSO/DSA91304A: 12.5 GHz
With DSO81004A: 10 GHz
With DSO/DSA91204A: 12 GHz
With DSO/DSA90804A: 8 GHz
Input Resistance
50
Ω
± 2%
Input Dynamic Range
± 1.1V Differential or Single-Ended
Maximum Input, (Vin – Vcm_term)
b
2.28 Vrms
Input Common Mode Range
± (4.3 V – Vcm_term x 0.29)
± 0.8 V
dc to 100 Hz
> 100 Hz
Maximum Signal Slew Rate
c
(SMA attenuator can extend range.)
d
25 V/ns
Differential Input
DC Attenuation
~2.2:1 (–6.9 dB)
Zero Offset Error Referred to Input
< 2 mV
Noise Referred to Input
1.6 mVrms (~ 14 nV/rtHz using noise BW of 12.5 GHz)
Propagation Delay
~6.15 ns
a Decreased rise and fall times mainly due to phase correction performed in the DSO80000 series, not due to DSP
boosting (except in DSO81304A).
b Vcm_term is the voltage supplied to the common mode termination port of the N5380B. If a shorting cap is in place, this
voltage is zero.
c SR max of sine wave = amplitude x 2 x p x frequency OR SR max of a step approximately equal to the amplitude x
0.6/trise (20-80%).
d Use of X:1 SMA coaxial attenuators in front of SMA probe head will: 1) Increase by X the max input signal slew rate,
dynamic range, offset range, common mode range, noise referred to the input, DC attenuation, and maximum input
voltage, 2) Most likely improve return loss or input TDR if attenuators are high quality, and 3) Not affect bandwidth and
rise time if attenuators are high quality.
Содержание 1168/9A
Страница 1: ...Agilent Technologies Agilent 1168 9A Differential and Single Ended Probes User s Guide ...
Страница 5: ...Contents 1168 9A Probes User s Guide 5 N5381A and N5382A Probe Heads 248 Other Accessories 249 ...
Страница 6: ...6 1168 9A Probes User s Guide Contents ...
Страница 9: ...Getting Started 1 Introduction 1168 9A Probe s User s Guide 9 Figure 2 Available Probe Heads ...
Страница 40: ...40 1168 9A Probe s User s Guide 1 Getting Started Service ...
Страница 69: ...Using Probe Heads 2 Recommended Configurations 1168 9A Probes User s Guide 69 Figure 36 Removing the Probe ...
Страница 104: ...104 1168 9A Probes User s Guide 2 Using Probe Heads N2887A 8A Soft Touch Probe Heads ...
Страница 120: ...120 1168 9A Probes User s Guide 3 Maintaining Probe Heads E2677A 9A Solder In Probe Heads ...
Страница 136: ...136 1168 9A Probes User s Guide 4 Calibrating Probes N2887A N2888A Calibration and Deskew Procedure ...
Страница 150: ...150 1168A 9A Probe s User s Guide 5 Characteristics and Specifications Probe Dimensions ...
Страница 202: ...202 1168A 9A Probe s User s Guide 6 Performance Plots N2887A N2888A Soft Touch Probe Heads ...
Страница 250: ...250 1168A 9A Probe s User s Guide 9 Replacement Parts Other Accessories ...