Getting Started
1
Using Offset With InfiniiMax Active Probes
1168/9A Probe’s User’s Guide
31
Using Offset With InfiniiMax Active Probes
It is important to understand how the 113xA probes behave
with respect to offset when different probe head / signal
combinations are used.
The purpose of offset in active probes or oscilloscope front
ends is to allow the subtraction of most or all of the dc
component of the input signal so the signal can better utilize
the dynamic range of the input. When using an InfiniiMax
probe with an Infiniium oscilloscope, you can select the case
(see the three cases described below) that applies for your
measurement by selecting the
Probes
button under the
channel setup menu. This allows you to select which type of
probe head is being used and, if it is a differential probe
head, allows you to select whether you are probing a
differential or single- ended signal. With these inputs, the
oscilloscope will use the proper type of offset for your
measurement case. The specifics for each case are discussed
below.
When adjusting the offset for a particular probe head, make
sure to have a triggered signal.
Case 1. A single-ended
probe head probing a
single-ended signal
For this case, the offset control on the oscilloscope controls
the probe offset and the channel offset is set to zero. This
allows the offset voltage to be subtracted from the input
signal before the signal gets to the differential amplifier.
Since this subtraction is done before any active circuits, the
offset range is large (±16V). Note that the minus probe tip is
not present when using a single- ended probe head which
means nothing is plugged into the "–" input of the probe
amp. This is normal and causes no problems.
Case 2. A differential probe
head probing a single-ended
signal
For this case, the offset control on the oscilloscope controls
the probe offset and the channel offset is set to zero. This
allows the offset voltage to be subtracted from the input
Содержание 1168/9A
Страница 1: ...Agilent Technologies Agilent 1168 9A Differential and Single Ended Probes User s Guide ...
Страница 5: ...Contents 1168 9A Probes User s Guide 5 N5381A and N5382A Probe Heads 248 Other Accessories 249 ...
Страница 6: ...6 1168 9A Probes User s Guide Contents ...
Страница 9: ...Getting Started 1 Introduction 1168 9A Probe s User s Guide 9 Figure 2 Available Probe Heads ...
Страница 40: ...40 1168 9A Probe s User s Guide 1 Getting Started Service ...
Страница 69: ...Using Probe Heads 2 Recommended Configurations 1168 9A Probes User s Guide 69 Figure 36 Removing the Probe ...
Страница 104: ...104 1168 9A Probes User s Guide 2 Using Probe Heads N2887A 8A Soft Touch Probe Heads ...
Страница 120: ...120 1168 9A Probes User s Guide 3 Maintaining Probe Heads E2677A 9A Solder In Probe Heads ...
Страница 136: ...136 1168 9A Probes User s Guide 4 Calibrating Probes N2887A N2888A Calibration and Deskew Procedure ...
Страница 150: ...150 1168A 9A Probe s User s Guide 5 Characteristics and Specifications Probe Dimensions ...
Страница 202: ...202 1168A 9A Probe s User s Guide 6 Performance Plots N2887A N2888A Soft Touch Probe Heads ...
Страница 250: ...250 1168A 9A Probe s User s Guide 9 Replacement Parts Other Accessories ...