DX-D 60C, DX-D 60G
Commercial name
DX-D 60C, DX-D 60G
Electrical connection DR Detector
Rated power supply
(powered by System Control Unit)
DC +24 V, Max. 1.0 A
Power consumption
max. 24 W
Environmental conditions (during normal operation)
Room temperature
b10 °C and +35 °C
Humidity (non condensing)
between 30% and 85% RH
(non-condensing)
Atmospheric pressure
between 700 hPa and 1060 hPa
Environmental conditions (during storage and transport)
Temperature (ambient)
between -15 °C and +55 °C
Humidity (non condensing)
between 10% and 90%
(non-condensing)
Atmospheric pressure
between 500 and 1060 hPa
Warming-up time
30 minutes
Dimensions
Dimensions
width x length x height
approx. 460 x 460 x 15.5 mm
Weight
4.2 kg
Maximum load
100 kg on an area of 40 mm in diameter
Maximum total load
150 kg over the whole detector surface
Vibration tolerance
2 G during normal operation
5G during storage and transport
Shock tolerance
20 G during normal operation
DX-D 60C, DX-D 60G | Technical Data |