![AGFA 5410/050 Скачать руководство пользователя страница 72](http://html.mh-extra.com/html/agfa/5410-050/5410-050_user-manual_2866605072.webp)
Corner area of detector not exposed
Details
A small area in the corner of the detector has not been ex-
posed.
Cause
When making an exposure with small source-image distance
(e.g. 1 m) and with the collimator in rotated position (e.g.
45°), the X-ray field does not reach to the outer corners of the
collimated area.
Solution
Increase the source-image distance.
72
| DX-D 100 | Problem solving
0187K EN 20191024 1609
Содержание 5410/050
Страница 1: ...DX D 100 5410 050 5411 050 5411 300 5411 400 User Manual 0187K EN 20191024 1609 ...
Страница 14: ...Figure 2 standard column Figure 3 telescopic column 14 DX D 100 Introduction 0187K EN 20191024 1609 ...
Страница 69: ...b remove the battery pack DX D 100 Getting started 69 0187K EN 20191024 1609 ...