Retrieving the patient info
Note
:
If you use a wired network the network cable must be plugged
in to be able to perform this step.
1.
On the RIS, schedule the examinations.
2.
In NX, query the RIS.
3.
Define the patient info for the examination.
4.
Start the examination.
Note
:
If the patient cannot be retrieved from the RIS you should
enter patient data manually. In that case, you can use the ‘virtual
keyboard’.
Related Links
on page 58
Selecting the exposure
In NX, select the thumbnail for the exposure in the Image Overview pane
of the Examination window.
The DR Detector is activated. The DR Detector Switch shows which DR
Detector is active and shows its status.
The DX-D 100 software console is displayed and the default X-ray exposure
parameters for the selected exposure are displayed.
Preparing the exposure
1.
Check the position of the X-ray unit and the patient.
WARNING:
Liquids ingressing the DR Detector may cause malfunction
and contamination.
If there is a chance that the detector comes in contact with liquids (bodily
fluids, disinfectants,...), the DR Detector must be wrapped in a protective
plastic bag while performing the examination.
• If a filter is used, mount the filter to the collimator.
• If a grid is used, attach the grid to the detector.
2.
Switch on the light localizer on the collimator. Adapt collimation if
required.
54
| DX-D 100 | Getting started
0187K EN 20191024 1609