5.
In a configuration with a wireless DR Detector, power on the DR Detector:
• attach a fully charged battery pack to the DR Detector.
• turn on the DR Detector.
• if needed, register the DR Detector to the DX-D 100 X-Ray System.
6.
At the start of each day and when the selected tube has not been in use for
approximately one hour, perform the X-ray tube warm-up as follows:
• Close the collimator blades fully.
• On the Control Console, display the software console by clicking the
SC
button.
• Select the free exposure workstation.
• Select 70 kV, 100 mAs, 200 mA and 500 ms exposure.
• Make sure that no one will be exposed.
• Make a total of three exposures, 15 seconds apart.
50
| DX-D 100 | Getting started
0187K EN 20191024 1609