Displacement and positioning workflow
CAUTION:
Before using DX-D 100, check the battery levels on the control
panel. If the red indicator is blinking, operation is not allowed.
In that case, it is necessary to charge the batteries.
Topics:
•
•
•
•
Positioning the X-ray tube and collimator
Driving the unit
Note
:
If you use a wired network the network cable must be
unplugged to be able to perform this step.
1.
Grip and hold the locking bar towards the handlebar.
1
2
1.
Locking bar
2.
Handlebar
2.
Push the handlebar with both hands to drive forwards.
Apply different pressure on the left and the right to steer the unit.
3.
Release the locking bar to block motion.
Releasing the arm
1.
Press the brake control at the tube-collimator handles to release the arm
from parking position.
2.
Press and hold the brake control while using the handles to rotate the
column and move the arm horizontally and vertically.
52
| DX-D 100 | Getting started
0187K EN 20191024 1609