![AGFA 5410/050 Скачать руководство пользователя страница 56](http://html.mh-extra.com/html/agfa/5410-050/5410-050_user-manual_2866605056.webp)
Executing the exposure
1.
Take a safe distance from the X-ray tube.
CAUTION:
Excessive user or operator radiation exposure. Always keep a
distance of at least 2 meters from the focal spot and X-ray
beam, protect body and do not expose hands, arms or other
parts of the body to the primary beam.
2.
On the handswitch or on the remote control, press the exposure button to
execute the exposure.
The exposure is sent to NX.
WARNING:
During exposure ionizing radiation is emitted by the X-ray
system. To indicate the presence of ionizing radiation, the
radiation indicator on the control console lights up.
CAUTION:
Do not select another thumbnail until the preview image is
visible in the active thumbnail. The acquired image may be
linked to the wrong exposure.
In NX:
• While the acquisition is ongoing, the X-ray on indicator is displayed on the
software console. The led beacon light lights up yellow.
• The image is acquired from the DR Detector and displayed in the
thumbnail.
• The software console disappears.
• If collimation is applied, the image is automatically cropped at the
collimation borders.
• The actual X-ray exposure parameters are shown in the Image Detail pane.
Performing a quality control
In NX:
1.
Select the image on which quality control is to be performed.
2.
Prepare the image for diagnosis by using e.g. L/R markers or annotations.
3.
If the image is OK, send the image to a hardcopy printer and/or PACS
(Picture Archiving and Communication System).
Note
:
If you use a wired network, NX can only send images if
the network cable is plugged in.
56
| DX-D 100 | Getting started
0187K EN 20191024 1609
Содержание 5410/050
Страница 1: ...DX D 100 5410 050 5411 050 5411 300 5411 400 User Manual 0187K EN 20191024 1609 ...
Страница 14: ...Figure 2 standard column Figure 3 telescopic column 14 DX D 100 Introduction 0187K EN 20191024 1609 ...
Страница 69: ...b remove the battery pack DX D 100 Getting started 69 0187K EN 20191024 1609 ...