Chapter 4 Maintenance Procedures
97
USG-CUS-059 Rev05
MassARRAY® Nanodispenser RS1000 v2.1 User Guide
For Research Use Only. Not for use in diagnostic procedures.
11. Close the main door.
12. Tap the
HOME
button in the
safety interlock is disengaged!
message.
Wait for the pin array assembly to move to home position (all the way left, forward,
and up).
Flush the calibrant
reservoir
Flush any calibrant from the calibrant reservoir that may remain from previous operation.
Even if calibrant will not be dispensed, flushing the calibrant reservoir is recommended to
keep it clean.
To flush the calibrant reservoir:
1.
Park the pin array assembly (tap the
PARK
button
on any screen).
Parking the pin array assembly allows easier access to the processing deck.
2.
Open the main door.
The
safety interlock is disengaged!
message appears.
3.
Remove the calibrant reservoir from the dry station by lifting it from the dry station as
shown in
. You may need to slightly rock the calibrant reservoir side-to-side
to lift it out.
Figure 4.11 Placing the Wash Supply Bottle on the Wash Station Block
Press
Ultrasonic wash
supply bottle
Wash station
block
Ultrasonic wash