Chapter 4 Maintenance Procedures
98
USG-CUS-059 Rev05
MassARRAY® Nanodispenser RS1000 v2.1 User Guide
For Research Use Only. Not for use in diagnostic procedures.
4.
Flush the calibrant reservoir with Type 1, >18.2
M
Ω
water.
5.
Dry the calibrant reservoir with clean dry air (for example, from a compressed air can).
CAUTION
!
Do not dry the reservoir with house air which can contain oil and other contaminants.
6.
Orient the calibrant reservoir as shown in
and insert it into the dry station,
making sure that it is firmly seated all the way down.
7.
Close the main door.
8.
Tap the
HOME
button in the
safety interlock is disengaged!
message.
Wait for the pin array assembly to move to home position (all the way left, forward,
and up).
Figure 4.12 Removing the Calibrant Reservoir
Figure 4.13 Calibrant Reservoir
Lift
Dry station
Calibrant
reservoir